Design and Prototyping of a Passive Spinal Brace with Modeled Nonlinear Resistance
Yeonsu Ahn
21/11/26
Passive spinal braces may be designed to distribute resistance differently across a movement interval. However, explicit mechanical comparisons may be required for such concepts. In this study, documented development of EMMA v1 was combined with mathematical analysis of a proposed nonlinear torque profile. A posterior core, lateral panels, and elastic ribs were incorporated by this prototype and fabricated using additive manufacturing. A linear-plus-quadratic model was compared with linear profiles that were matched either for initial stiffness or for terminal torque. Evaluating three illustrative nonlinear parameter settings over 0 - 30°, a transition was assumed at 10°. Relative to linear profiles matched at 30°, the nonlinear profiles reduced initial stiffness by 40.0%, 57.1%, and 72.7% and loading work by 22.2%, 31.7%, and 40.4%, respectively. Greater terminal resistance was provided by nonlinear models against profiles matched for initial stiffness, requiring more work. During prototype development, hinge-tab thickness increased by 14.3% and rib amplitude decreased by 33.3%; these changes were not validated through instrumented mechanical testing. Conditional tradeoffs were established by the findings within the model providing quantitative benchmarks for evaluating future prototypes. To determine whether the fabricated device can realize these resistance profiles or support postoperative rehabilitation, model calibration and physical testing remain necessary.